technical Data
MiniCal II
Electrodes:
Ion-selective measurement of
- Ammonia (NH3)
- Ammonium (NH4)
- Reference-electrodes
- Freon, FHC, FCHC
- Redox - mV - Measurement
- Oxygen measurement
Detection ranges:
NH3 and NH4:
0..10 / 0...100 ppm
FHC/FCHC:
0..100 ppm / 0..1000 ppm
Oxygen in water:
0.05 mg/l .. 10 mg/l dissolved
oxygen under normal conditions
Detection limit:
From 0.2 ppm for NH4
Voltage supply:
24 V DC ±20%
Power consumption:
Max. 55 mA
Output signal:
4 .. 20 mA
for 4 measurement signals
indirect coupled
max. load 100 ohm
Casing:
Chrome-steel, for outside mounting
Splash water proof IP 65
Operation
2 button operation for settings and adjustment
Temperature compensation:
NTC
Cooling water probe
Pressure range:
0.... 6 bar
Temperature Measurement-medium:
-35 .. 80 °C (W12051.II)
-35 .. 120 °C (W12051.II/H)
(accompanying heater < -25° C)
Medium connecting parts:
POM (W12051.II) bis 80°C
PVDF (W12051.II/H) > 80°C
Mounting:
DIN-flange DN50, PN16
(see mounting instructions)
Weight:
ca. 2 Kg
|
|